Setup is different for each model. Please click on link below and refer to the respective user manual/quick start guide:
Product : product.php
Download : download.php
Setup is different for each model. Please click on link below and refer to the respective user manual/quick start guide:
Product : product.php
Download : download.php
Each meter is designed to admit specifically-designed test strips of its own. Hence, please note test strips for 1) other models and 2) other brands will not work.
Expiry dates are printed on the strip box.
Depending on the type of test strips you have on hand, they may be preserved from 24 to 18 months. To ensure the accuracy of test results, please note test strips should be used within 3 months from first opening.
Common reasons that may cause test strips to become damp:
1.If you use wet hands to take out test strips
2.If you do not close test vial instantly after you take out.
3 If you do not storage the test in specification way.
1. Please store the test strips at room temperature between 4-30℃(39-86℉). Do not put them in the fridge.
2. Please use the test strips at temperatures between 10℃(50℉) and 40℃(104℉),and less than 85% humidity.
3. Please keep the test strips in their original vials only; do not transfer them into another vial or container.
4. Always remember to close the cap of vials immediately after use.
5. Avoid exposure of test strips to air for a long time.
1. To avoid infection, lancets can only be used once; do not reuse it.
2. The needles on the lancets are sterilized; do not wipe the needles with alcohol pads before use.
3. When using lancets, please keep the protective cover on. After testing, put the protective cover back to the lancet.
Use an alcohol pad to sterilize your finger first and make sure your finger is dry before pricking your fingertip.
1. According to EN ISO15197, the accuracy of the blood glucose meter is determined by comparing with standard biochemical analyzer and not other blood glucose meters.
2. According to EN ISO15197:2015 standard, the variation between the indices of blood glucose meters and the data of the reference standard biochemical analyzer is subject to the below restrictions:
Please reach out to our local distributor (where you purchased our products) or contact us directly via link below:
Please contact our authorized distributor in your area.
The meter test range is 20~600mg/dL (1.1~33.3mmol/L). Results that are out of this range will result in error message as below:
For test results greater than 600mg/dL (33.3mmol/L), the meter will display 「HI」.
For test results lower than 20mg/dL(1.1mmol/L), the meter will display 「LO」.
(For eBchek, eBwell, eBfresh, eBcare)
(For eBsensor, eBcarryon, eBfine)
When the meter displays “HI”or “LO”, please retake the test. If the result pertains, please contact your doctor immediately.
When the meter displays the error message below, it is out of power. Please replace with a new battery.
1. The symbol of battery is flashing.
2. The symbol of battery and LO are flashing.
(For eBchek, eBwell, eBfresh, eBcare)
(For eBsensor, eBcarryon, eBfine)
If "E01" mark is display, it indicates that the meter is malfunction. Please remove the test strip and put the meter back to the package, and reach out to our local distributor.
(For eBchek, eBwell, eBfresh, eBcare, eBsensor, eBcarryon, eBfine)
(For eBchek, eBwell, eBfresh, eBcare)
This happens when the surrounding temperature goes above 42℃(107.6℉) or below 4℃(39.2℉). When this happens the error screen below will appear and the meter will shut off automatically. Please move the meter to an ambient place 10℃-40℃(50℉~104℉) for 20 minutes and retake the blood glucose test.
(For eBsensor, eBcarryon, eBfine)
This happens when the surrounding temperature goes above 42℃(107.6℉) or below 4℃(39.2℉). The meter displays "HI" for the former and "LO" for the latter. When this happens, the error screen below will appear and the meter will shut off automatically.
Please move the meter to an ambient place 10℃~40℃(50~104℉) for 20 minutes and retake the test. If the issue persists, please reach out to our local distributor.
When the "E-U" error message is shown on the screen, there are two possibilities: either 1) the operating steps are done incorrectly or 2) the test strip has previously been used. Kindly note each test strip can only be used once, and only once. Please replace with a new test strip and retake the test. If the issue persists, please reach out to our local distributor.
(For eBchek, eBwell, eBfresh, eBcare, eBsensor, eBcarryon, eBfine)
When the "HI" error message is displayed, there are two possibilities: either 1) the testing result is greater than 8.0mmol/L and therefore out of the meter's range or 2) the test strip is malfunctioning. For the latter scenario, please replace with a new test strip and test again. If the issue persists, please reach out to our local distributor.
When the "E01" error message is displayed, the meter is malfunctioning. Please remove the test strip, put the meter back to the package and reach out directly to our local distributor. If the issue persists, please reach out to our local distributor.
When the "E03" error message is displayed, you have inserted the wrong test strips (most likely test strips from another brand). Please replace the test strips with ours and retake the test. If the issue persists, please reach out to our local distributor.
(eBketone, eBketolife)
This happens when the surrounding temperature goes above 42℃ (107.6℉) or below 4℃(39.2℉). When this happens, the error screen below will appear and the meter will shut off automatically.
1. A symbol of thermometer
2. A symbol of thermometer and " LO"
3. A symbol of thermometer and "HI"
Please move the meter to an ambient place of 10℃~40℃ (50℉~104℉) for 20 minutes and retake the test. If the issue persists, please reach out to our local authorized distributor.
(eBketoQ, eBketoway)
This happens when the surrounding temperature goes above 42℃(107.6℉) or below 4℃(39.2℉). When this happens, the error screen below will appear and the meter will shut off automatically. Please move the meter to an ambient place 10℃-40℃ (50℉~104℉) for 20 minutes and retake the test. If the issue persists, please reach out to our local distributor.
(eBketone, eBketolife)
This indicates that the test strip has been used or damp. Please replace it with a new test strip and retake the test again. If the issue persists, please reach out to our local distributor.
(eBketoQ, eBketoway)
This indicates that the test strip is used or damp. Please replace it with a new test strip and retake it again. If the issue persists, please reach out to our local distributor.
(eBketone, eBketolife)
This indicates that either 1) the code card is damaged or 2) the wrong code card is used. Please check if the code on the code card matches with that of the test strip vial. If the issue persists, please reach out to our local distributor.
When the meter displays the error message below, it is out of power. Please replace it with a new battery.
1. The symbol of the battery
2. The symbol of battery and "LO"
(eBketoQ, eBketoway)
When the meter displays the error message below, it is out of power and will shut off automatically. Please replace it with a new battery.
When a blood drop mark and the "HI" error message are displayed at the same time, there are two possibilities: either 1) the testing result is greater than 20 mg/dL (1190 μmol/L) and therefore out of the meter's range or 2) the test strip is malfunctioning. For the latter scenario, please replace it with a new test strip and retake the test again. If the issue persists, please reach out to our local distributor.
When the "E01" error message is displayed, the meter is malfunctioning. Please remove the test strip, put the meter back in the package, and reach out directly to our local distributor.
When the "E03" error message is displayed, you have inserted the wrong test strips (most likely test strips from another brand). Please replace the test strip with ours and retake the test. If the issue persists, please reach out to our local distributor.
This happens when the surrounding temperature goes above 42℃ (107.6℉) or below 4℃ (39.2℉). When this happens, the error screen below will appear and the meter will shut off automatically.
1. A symbol of thermometer
2. A symbol of thermometer and " LO"
3. A symbol of thermometer and "HI"
Please move the meter to an ambient place of 10℃~40℃ (50℉~104℉) for 20 minutes and retake the test. If the issue persists, please reach out to our local distributor.
This indicates that the test strip has been used or is damp. Please replace it with a new test strip and retake the test again. If the issue persists, please reach out to our local distributor.
This indicates that either 1) the code card is damaged or 2) the wrong code card is used. Please check if the code on the code card matches with that of the test strip vial. If the issue persists, please reach out to our local distributor.
When the meter displays the error message below, it is out of power. Please replace with a new battery.
1. The symbol of battery
2. The symbol of battery and "LO"
When a blood drop mark and "HI" error message are displayed at the same time, there are two possibilities: either 1) the test result is greater than 20mg/dL (1190μmol/L ) and therefore out of the meter's range or 2) the test strip is malfunctioning. For the latter scenario, please replace it with a new test strip and retake the test again. If the issue persists, please reach out to our local distributor.
When the "E01" error message is displayed, the meter is malfunctioning. Please remove the test strip, put the meter back in the package, and reach out directly to our local distributor.
When the "E03" error message is displayed, you have inserted the wrong test strips (most likely test strips from another brand). Please replace the test strip with ours and retake the test. If the issue persists, please reach out to our local distributor.
This happens when the surrounding temperature goes above 42℃ (107.6℉) or below 4℃ (39.2℉). When this happens, the error screen below will appear and the meter will shut off automatically.
1. A symbol of thermometer
2. A symbol of thermometer and "HI"
3. A symbol of thermometer and " LO"
Please move the meter to an ambient place 10℃~40℃(50℉~104℉) for 20 minutes and retake the test. If the issue persists, please reach out to our local distributor.
This indicates that the test strip has been used or is damp. Please replace with a new test strip and retake the test again. If the issue persists, please reach out to our local distributor.
This indicates that either 1) the code card is damaged or 2)the wrong code card is used. Please check if the code on the code card matched that of the test strip vial. If the issue persists, please reach out to our local distributor.
When the meter displays the error message below, it is out of power. Please replace it with a new battery.
1. The symbol of battery.
2. The symbol of battery and "LO".